Viewer

org.bluetooth.characteristic.plx_features

Last Modified: 2015-07-14 Approved: Yes

Name: PLX Features

Type: org.bluetooth.characteristic.plx_featuresDownload / View

Assigned Number: 0x2A60

Summary:

The PLX Features characteristic is used to describe the supported features of the Server. Included in the characteristic is a PLX Features field, and, depending on the contents of the PLX Features field, the Measurement Status Support field, and the Device and Sensor Status Support field.

Value Fields

Names Field Requirement Format Minimum Value Maximum Value Additional Information
Supported Features
Mandatory
16bit N/A N/A
Bit Field
Bit Size Name
Definition
Key Value Requires
0 1 Measurement Status support is present
0 False
1 True C1
1 1 Device and Sensor Status support is present
0 False
1 True C2
2 1 Measurement Storage for Spot-check measurements is supported
0 False
1 True
3 1 Timestamp for Spot-check measurements is supported
0 False
1 True
4 1 SpO2PR-Fast metric is supported
0 False
1 True
5 1 SpO2PR-Slow metric is supported
0 False
1 True
6 1 Pulse Amplitude Index field is supported
0 False
1 True
7 1 Multiple Bonds Supported
0 False
1 True
8 8 Reserved for future use
Measurement Status Support
C1
16bit N/A N/A
Bit Field
Bit Size Name
Definition
Key Value
5 1 Measurement Ongoing bit supported
0 False
1 True
6 1 Early Estimated Data bit supported
0 False
1 True
7 1 Validated Data bit supported
0 False
1 True
8 1 Fully Qualified Data bit supported
0 False
1 True
9 1 Data from Measurement Storage bit supported
0 False
1 True
10 1 Data for Demonstration bit supported
0 False
1 True
11 1 Data for Testing bit supported
0 False
1 True
12 1 Calibration Ongoing bit supported
0 False
1 True
13 1 Measurement Unavailable bit supported
0 False
1 True
14 1 Questionable Measurement Detected bit supported
0 False
1 True
15 1 Invalid Measurement Detected bit supported
0 False
1 True
0 5 Reserved for future use
Device and Sensor Status Support
C2
24bit N/A N/A
Bit Field
Bit Size Name
Definition
Key Value
0 1 Extended Display Update Ongoing bit supported
0 False
1 True
1 1 Equipment Malfunction Detected bit supported
0 False
1 True
2 1 Signal Processing Irregularity Detected bit supported
0 False
1 True
3 1 Inadequite Signal Detected bit supported
0 False
1 True
4 1 Poor Signal Detected bit supported
0 False
1 True
5 1 Low Perfusion Detected bit supported
0 False
1 True
6 1 Erratic Signal Detected bit supported
0 False
1 True
7 1 Nonpulsatile Signal Detected bit supported
0 False
1 True
8 1 Questionable Pulse Detected bit supported
0 False
1 True
9 1 Signal Analysis Ongoing bit supported
0 False
1 True
10 1 Sensor Interface Detected bit supported
0 False
1 True
11 1 Sensor Unconnected to User bit supported
0 False
1 True
12 1 Unknown Sensor Connected bit supported
0 False
1 True
13 1 Sensor Displaced bit supported
0 False
1 True
14 1 Sensor Malfunctioning bit supported
0 False
1 True
15 1 Sensor Disconnected bit supported
0 False
1 True
16 8 Reserved for future use

Note: